Developments in Dental Implant Surface Modifications – a Review
Department of Prosthodontics & Implant Dentistry, Mamata Dental College, Khammam
ABSTRACT
Dental implant success depends on achieving rapid, stable osseointegration. Surface properties, topography and chemistry significantly influence bone healing and long-term outcomes. Implant surfaces have evolved from smooth, machined titanium to complex textured and bioactive coatings. Moderately rough surfaces (Sa ~1–2 μm) typically enhance bone–implant contact and healing. Modification methods include subtractive techniques (acid etching, grit-blasting, laser ablation) and additive/coating methods (anodization, plasma spraying, biomimetic deposition). Each alters micro/nano-scale features affecting osteoblast behavior, soft-tissue response, and bacterial adhesion. Recent advances involve nanoscale engineering, bioactive coatings, drug delivery systems, and alternative materials like zirconia. This review summarizes the evolution of implant surfaces, classification by topographic scale, key modification techniques, biological responses, implications for peri-implant diseases, and emerging trends such as multifunctional smart coatings. Clinical implications are highlighted to guide future implant design and patient care.
Keywords: dental implants; surface modification; osseointegration; surface roughness; biomimetic coating; zirconia
INTRODUCTION
Dental implants have transformed tooth replacement, with 10-year survival rates of 95--98%.1,2 Success hinges on osseointegration, a direct, functional bond between bone and implant.1 While many factors influence this process, surface properties play a critical role: optimized roughness and chemistry accelerate healing and enhance mechanical interlock.3,4,5 Albrektsson et al. defined implant success by immobility and absence of infection.1 Surface modifications aim to improve outcomes in poor bone or systemic illness.4 Micro-rough surfaces promote osteoblast adhesion and differentiation, translating clinically to faster healing and greater early stability.5,6
Implant surface engineering now spans simple mechanical texturing to nano-coatings and drug-delivery systems.7,8 This review outlines the historical evolution of implant surfaces, categorizes them by macro, micro, and nano-topography, and details major modification techniques, both subtractive and additive.2,5,9 Their biological effects on osseointegration and peri-implant tissues are discussed.6,10 Finally, we examine emerging innovations like bioactive coatings, nanotechnology, drug delivery, and alternative materials such as zirconia.11,12,13 Clinical advantages and limitations of each approach are considered, with an emphasis on current evidence and future directions.2,8
HISTORICAL OVERVIEW
The concept of osseointegration in dentistry began with Brånemark et al. in the late 1960s using threaded, smooth (machined) titanium implants. These first-generation implants had turned surfaces with minimal roughness. Healing protocols required 3-6 months of unloaded integration due to slow bone response. In the 1970s-80s, no surface treatments were added beyond cleaning.14
In the late 1980s-1990s, surface roughening emerged to enhance bone contact and reduce healing time. Techniques like grit-blasting and acid etching gained popularity. For instance, Straumann's SLA surface (~1.5 µm Sa) used alumina blasting followed by acid etching.15,16 Zimmer's MTX surface used calcium phosphate media, which dissolved, leaving a moderately rough oxide surface.17 These second-generation surfaces doubled bone-implant contact, improving stability.5,6 Animal studies confirmed stronger bone anchorage with blasted surfaces.18.
Additive coatings soon followed. In the 1990s, hydroxyapatite (HA) and titanium plasma spray (TPS) coatings were introduced, yielding very rough surfaces (Ra ~5-7 µm).19,20 Though HA initially promoted bone growth, issues like delamination and inflammation limited long-term success.21,22,23 Moderately rough surfaces (Sa ~1-2 µm) later became standard.6,15 SLActive (a hydrophilic variant of SLA) and anodized surfaces like TiUnite (Nobel Biocare) further accelerated healing.24,25 Ceramic (zirconia) implants emerged as a metal-free option, later modified with micro-roughening to enhance performance.11,12,13
In summary, implant surfaces evolved from smooth, machined titanium (first generation) to roughened, bioactive coatings (second/third generation), reducing healing time and improving outcomes in poor bone.6,18 Concerns about bacterial colonization and coating longevity have led to current research on multifunctional coatings.26,27
CLASSIFICATION OF SURFACE MODIFICATIONS
Implant surfaces are classified by feature size and morphology.
Macrotopography refers to the overall implant shape, threads, flutes and cavities which affect primary stability and load distribution.3 Thread design (e.g., square, V-shaped) influences stress transfer to bone. These features, visible to the naked eye, are optimized for clinical scenarios (e.g., tapered bodies for extraction sockets).4
Microtopography includes surface roughness in the micrometer range (Sa ~0.5-4 µm), crucial for osseointegration.5,6,28 Surfaces are categorized as minimally (0.5-1 µm), moderately (1-2 µm), or very rough (>2 µm).2,6 Moderately rough surfaces enhance bone-implant contact and integration.5,6 Microroughness is created by machining, blasting, or etching, producing features that mimic resorption lacunae (~5-10 µm).29 Most implants combine macrothreads for stability and microtexture for cellular attachment.3,4
Nanotopography involves features <100 nm.8 As natural bone ECM contains nanoscale irregularities, engineered nano-roughness aims to replicate this environment.8,30 Techniques like anodization or acid etching create nanopores/ridges.25,31 Advanced methods add nanocoatings or particles to influence protein adsorption and osteogenic behavior.32,33 Anodized titanium surfaces with 30-300 nm pores stimulate bone cell proliferation.25 Hierarchical surfaces combining micro- and nano-roughness provide both mechanical interlock and enhanced cell signaling.32
Surfaces are also classified by material or chemistry. Examples include: HA-coated (calcium phosphate)19, biomimetic (precipitated CaP)34, anodized (thickened oxide)25, and biochemical (growth factors, antibiotics).35 These chemical treatments often complement physical roughness to further improve biological response.19,34
TECHNIQUES FOR SURFACE MODIFICATION
Implant surfaces are modified via subtractive or additive processes (see Table 1). Subtractive methods remove material to create roughness, while additive methods deposit coatings.8,36
SUBTRACTIVE TECHNIQUES
Acid Etching
Titanium is treated with acids (HCl, H2SO4, HF), creating micro-pits (~0.5-2 µm).8,37 This sandpaper-like texture improves osteoblast retention and migration.38 The BIOMET 3i Osseotite uses dual acid-etching to yield ~1 µm microroughness.8 While etching alone gives modest roughness, it's commonly combined with blasting for enhanced results.9,15
Grit-Blasting (Sandblasting)
Hard particles (Al2O3, TiO2, HA) are blasted onto the implant, producing cratered surfaces.9,15,18 Alumina particles (~25--75 µm) yield Ra ~0.5-1.5 µm; larger particles create rougher profiles.15,18 TiO2-blasted surfaces show better bone anchorage than machined ones. Resorbable Blast Media (RBM) uses bioresorbable ceramics like calcium phosphate to leave no residue.17
Laser Ablation/Texturing
Pulsed lasers (Nd:YAG, Er:YAG) create micro-channels or nodular craters.39,40 Laser-treated surfaces orient osteoblast attachment and increase surface purity and hardness. BioHorizons' Laser-Lok uses 8-12 µm microgrooves to promote bone and soft-tissue integration.41
Other Subtractive Methods
Electropolishing yields glossy, smooth surfaces (Sa <0.4 µm), now rarely used.42 UV Photo functionalization modifies titanium oxide chemistry, increasing hydrophilicity and potentially enhancing early healing though clinical data is still emerging.24
ADDITIVE TECHNIQUES
Titanium Plasma Spraying (TPS)
Melted titanium is sprayed to form thick, rough coatings (Ra ~5-7 µm). While enhancing bone interlock, risks include delamination and titanium debris inflammation.20 TPS has largely been replaced by thinner, more stable coatings.8,36
Hydroxyapatite (HA) Coatings
HA, mimicking bone mineral, is plasma-sprayed or applied via other techniques1.9,43 Early HA coatings (<20 µm) boosted bone growth.19,20 However, they can dissolve or delaminate, causing inflammation.21,22,23 Modern systems favor thin, crystalline CaP layers (via sol-gel, sputtering) for better adhesion.43
Anodization (Micro-arc Oxidation)
Electrochemical anodizing thickens titanium oxide to several microns, producing nano-porous surfaces (1-2 µm pores).25 TiUnite is a classic anodized surface with strong bone integration.25 While bioactive, anodized surfaces may still require blasting for adequate microroughness.8,24
Biomimetic Calcium-Phosphate (CaP) Deposition
These coatings are applied in near-body temperature conditions (e.g., simulated body fluid immersion).34,44,45 CaP coatings (~1-5 µm) release Ca2+ and PO4 3- ions, stimulating osteoblasts.44,45 They may include BMP-2, collagen, or drugs.35 While promising, thick or poorly bonded coatings risk delamination.36
Sputtering and Laser Deposition
These apply ultra-thin coatings (TiO2, ZrO2, HA) with precise control. Sputtered layers (~0.5-3 µm) are dense but often amorphous.43 Laser deposition produces crystalline films, though it's costly.39 Doped coatings (e.g., silver for antimicrobial activity) are being explored.46
Electrophoretic Deposition (EPD)
Nano-HA particles are deposited using an electric field to form thin coatings. While effective for complex shapes, sintering may induce cracks due to thermal mismatch.45
Table 1. Summary of Common Surface Modification Techniques
Technique |
Example |
Roughness (Sa) |
Key Features |
Advantages |
Limitations |
|
Machined (turned) |
Brånemark |
~0.5 µm |
Baseline smooth surface |
Clean |
Less bone interlock |
|
Acid-etched |
3i Osseotite |
0.5–1 µm |
Micro-pits; improved osteoblast retention |
Uniform, clean |
Less rough than blasting |
|
Grit-blasted |
SLA (TiO2) |
1–2 µm |
Craters increase surface area |
Simple, effective |
Residue risk (Al2O3) |
|
RBM |
Zimmer MTX |
~1.5 µm |
Bioactive, dissolvable media |
No inert debris |
Needs precision |
|
SLA (Blast + Acid) |
Straumann |
~1.5 µm |
Dual roughening; high BIC |
Optimized profile |
Multi-step, irreversible |
|
Anodized |
TiUnite |
1–2µm (oxide) |
Porous oxide; high wettability |
Nanotopography |
May need blasting |
|
Laser-modified |
Laser-Lok |
~8µm grooves |
Oriented micro-channels |
Soft/hard tissue integration |
High cost |
|
TPS |
Straumann ITI |
~5–7 µm |
Thick Ti coating |
Bone interlock |
Delamination, debris |
|
HA Coating |
Plasma-sprayed |
<20 µm |
Osteoconductive layer |
Rapid bone growth |
Dissolution, inflammation |
|
Biomimetic CaP |
Dip/Immersion |
<5 µm |
Nanocrystalline; bioactive |
Low-temp |
Weak bond if thick |
|
Sputtered Films |
HA, TiO2 |
0.1–3 µm |
Uniform, thin |
High adhesion |
Amorphous, expensive |
|
Sol-gel CaP |
Dip + heat |
0.1–2 µm |
Thin glassy CaP |
Cost-effective |
Brittle |
|
Ion Beam |
Hybrid coating |
<30 nm |
Ultra-thin, precise |
Strong bond |
Research-stage |
|
EPD |
Nano-HA |
~1–5 µm |
Porous, uniform |
Coats complex shapes |
Cracking risk with sintering |
BIOLOGICAL RESPONSES TO IMPLANT SURFACES
After implant placement, healing begins with blood clot formation, inflammation, and bone deposition. The implant surface mediates this process. Initially, blood proteins (e.g., fibronectin, vitronectin) adsorb to the surface.47 Microrough textures entrap the fibrin clot, aiding osteogenic cell migration, while smoother surfaces may shed the clot more easily.9,29 Thus, moderately rough surfaces better retain the clot, supporting early osseointegration.5,6
Osteoprogenitor cells migrate along this fibrin matrix, differentiate into osteoblasts, and begin bone matrix deposition.47 Grooves and pits (~10 µm) mimic osteoclast resorption lacunae, guiding osteoblast alignment and natural bone deposition.29,41
Surface roughness also influences cell signaling.38,26 On micro-rough titanium, osteoblasts show elevated alkaline phosphatase activity and collagen production26. Nanotopography further enhances protein adsorption and integrin-mediated adhesion.32,33 Combining micro- and nano-roughness (hierarchical surfaces) maximizes osteoblast proliferation and differentiation.32
Different cell types respond differently to surface topography.29,48 Fibroblasts and epithelial cells prefer smoother textures, important for soft tissue integration. Hence, many implants feature a roughened body (for bone) and a smooth collar (for gingiva) to optimize both bone bonding and mucosal seal.
Wettability and surface energy also affect healing.24 Hydrophilic surfaces (e.g., SLActive) enhance protein adsorption and osteoblast adhesion, accelerating bone formation.24,25 Chemically modified SLA surfaces improve early healing by increasing surface energy and fibrin binding.24
In summary, implant surface properties, topography and chemistry govern protein adsorption, cellular attachment, and bone regeneration.32,35 Moderately rough, hydrophilic surfaces consistently show greater and faster bone-to-implant contact in both animal and human studies.15,34
Role of Surface Modifications in Peri-implant Diseases
While rough surfaces enhance osseointegration, they may also increase the risk of peri-implant disease. Rougher surfaces promote bacterial adhesion and biofilm formation. Studies show biofilm accumulation rises with surface roughness (Ra), up to a certain threshold. Surface free energy and hydrophobicity also influence colonization. Hydrophobic, rough surfaces may favor pathogenic microbes.26,49,50,51
Peri-implant mucositis and peri-implantitis, marked by inflammation and bone loss often stem from microbial colonization of exposed implant surfaces.52 When bone recedes, rough or coated surfaces become accessible to bacteria.53,54 HA coatings may be especially prone to bacterial adhesion compared to polished titanium.46 Porous or irregular coatings can harbor anaerobes, making decontamination difficult.55,56
In contrast, smoother implant collars may facilitate hygiene and reduce mucositis risk.51 Some designs include machined necks or laser-microgrooved zones to promote soft-tissue attachment while limiting microbial ingress.29,41
Clinically, long-term implant success depends not just on bone integration but on sustained peri-implant health. Surface design must balance osseointegration benefits with the risk of biofilm development.26 Hybrid designs featuring a roughened subcrestal portion and smooth supracrestal collar, aim to optimize this balance.34 In parallel, antimicrobial surface strategies are being developed to reduce peri-implantitis risk on rough surfaces.46
Emerging Trends
Implant surface research is moving beyond passive roughness toward smart, multifunctional designs.
Nanotechnology and Nano-engineering
Nanoscale modifications aim to fine-tune cellular responses. Titanium dioxide nanotube arrays (via anodization) can store and release drugs or peptides. Graphene or carbon coatings add strength and bioactivity. Techniques like electron-beam lithography create bone-mimetic nanopatterns that enhance gene expression and bone formation. Combining nano- and micro-roughness optimizes osseointegration.8,32
Bioactive and Biomimetic Coatings
"Smart" coatings deliver biological cues. These include growth factors (BMP-2, PDGF), peptides (RGD, antimicrobial), and ECM proteins (fibronectin, laminin).35,57,58,59 RGD peptides or fibrinogen fragments covalently bound to titanium increase osteoblast attachment.58 Calcium phosphate coatings gradually dissolve, releasing ions that stimulate bone mineralization.34,44 A recent review supports biomolecular coatings for enhancing early bone formation. Challenges include cost and stabilizing biologics on surfaces.35,58
Drug-Delivery Surfaces
Implants are being designed to release antibiotics (e.g., gentamicin-HA coatings), bisphosphonates, or osteoinductive agents like statins. Smart coatings may respond to local pH (infection) and release antimicrobials as needed. Though promising in animals, precise release control and clinical validation are needed.8,35
Peptide and Protein Engineering
Synthetic peptides mimicking bone proteins (e.g., osteopontin-derived) have been grafted onto titanium to enhance bonding. Some carry anti-inflammatory functions. Reviews suggest these coatings enhance osseointegration and soft-tissue seal. However, protein stability and clinical studies remain hurdles.58,59
Alternative Materials - Zirconia
Zirconia (Y-TZP) implants are strong, esthetic, and metal-free, but inherently less osteoconductive. Surface modification is essential. Sandblasting improves osteoblast adhesion by up to 175%. Acid etching alone is insufficient; combining with sandblasting (SLA-like) improves results.11,12 However, sandblasting may cause low-temperature degradation (LTD), weakening zirconia. New approaches such as laser microtexturing, glass or CaP coatings are under study.43
Additive Manufacturing (3D Printing)
Electron beam melting and laser sintering produce implants with controlled porosity and macro/nanotopography. Researchers are applying nanostructured coatings to 3D-printed implants for personalized designs. These approaches allow tailored roughness and drug-loading capacity.8
Antibacterial/Biofilm-Resistant Surfaces
Peri-implantitis has spurred the development of antibacterial surfaces. Strategies include superhydrophobic and zwitterionic polymers, or coatings containing silver or copper nanoparticles. Metal-organic frameworks (MOFs) and antimicrobial peptides are also under exploration. The challenge remains ensuring these strategies resist bacteria without impairing osseointegration.46,59
In summary, surface innovations are converging toward multifunctionality: combining optimal roughness with bioactivity, antimicrobial resistance, and possibly patient-specific responses. Integration of physical, chemical, and biological design holds promise for enhancing implant longevity.
CONCLUSION
Implant surface modifications have progressed from smooth titanium screws to multiscale engineered designs. Today’s surfaces integrate macro-thread architecture, microroughness (Sa ~1–2 µm), and, increasingly, nanoscale features to optimize osseointegration. Subtractive techniques like grit-blasting and acid etching remain foundational, while additive approaches such as anodization and thin bioactive coatings further enhance biological response.
Emerging strategies now aim to create bioactive, multifunctional surfaces. Coatings that deliver growth factors, antimicrobial agents, or drugs show promise, as does nano-engineering for fine control of cellular behavior. Zirconia and 3D-printed implants expand the design palette, supporting custom geometries and material profiles.
Clinically, implant surface choice should reflect patient-specific factors. Rough surfaces offer faster early stability and are useful in low-density bone but require meticulous hygiene due to increased biofilm risk. Coated surfaces may speed healing but can raise concerns about long-term durability, especially with thicker or poorly bonded layers. Long-term data supports success with moderately rough surfaces (e.g., SLA), while newer technologies (e.g., peptide-coated, laser-textured) await broader validation.
Dental implant surface science now bridges materials engineering and biology. No single surface suits all situations; future directions point toward tailored, patient-specific designs. A sound understanding of available surface modifications and their biological implications remains essential for maximizing clinical outcomes.
REFERENCES
- Albrektsson T, Zarb G, Worthington P, Eriksson AR. The long-term efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Implants. 1986;1:11-25.
- Smeets R, Stadlinger B, Schwarz F, et al. Impact of dental implant surface modifications on osseointegration. Biomed Res Int. 2016;2:1-15.
- Boyan B, Dean D, Lohmann C. The titanium-bone cell interface in vitro: the role of the surface in promoting osseointegration. In: Brunette D, et al., ed. Titanium in Medicine. Springer; 2001.
- Wennerberg A, Albrektsson T. Suggested guidelines for the topographic evaluation of implant surfaces. Int J Oral Maxillofac Implants. 2000;15:331-344.
- Wennerberg A. Albrektsson T. Effects of titanium surface topography on bone integration: a systematic review. Clin Oral Implants Res. 2009:20:172-184.
- Le Guéhennec L, Soueidan A, Layrolle P, Amouriq Y. Surface treatments of titanium dental implants for rapid osseointegration. Dent Mater. 2007;23:844-854.
- Rasouli R, Barhoum A, Uludag H. A review of nanostructured surfaces and materials for dental implants: surface coating, patterning and functionalization for improved performance. Biomater Sci. 2018;6:1312-1338.
- Dohan Ehrenfest DM, Coelho PG, Kang BS, et al. Classification of osseointegrated implant surfaces: materials, chemistry and topography. Trends Biotechnol. 2010;28:198-206.
- Alla RK, Ginjupalli K, Upadhya N, et al. Surface roughness of implants: a review. Trends Biomater Artif Organs. 2011;25:112-118.
- Mendonça G, Mendonça DBS, Aragão FJL, Cooper LF. Advancing dental implant surface technology - From micron- to nanotopography. Biomaterials 2008:29:3822-3835.
- Oliva J, Oliva X, Oliva JD. Five-year success rate of 831 consecutively placed Zirconia dental implants in humans: a comparison of three different rough surfaces. Int J Oral Maxillofac Implants. 2010;25:336-344.
- Özkurt Z, Kazazoglu E. Zirconia dental implants: a literature review. J Oral Implantol. 2011;37:367-376.
- Gahlert M, Gudehus T, Eichhorn S, et al. Biomechanical and histomorphometric comparison between zirconia implants with varying surface textures and a titanium implant in the maxilla of miniature pigs. Clin Oral Implants Res. 2007;18:662-668.
- Buser D, Schenk RK, Steinemann S, et al. Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs. J Biomed Mater Res. 1991;25:889-902.
- Fischer K, Stenberg T. Prospective 10-year cohort study based on a Randomized Controlled Trial (RCT) on implant-supported Full-Arch Maxillary Prostheses. Part 1: sandblasted and acid-etched implants and mucosal tissue. Clin Implant Dent Relat Res. 2012;14:808-815.
- Wennerberg A, Albrektsson T, Lausmaa J. Torque and histomorphometric evaluation of c.p. titanium screws blasted with 25- and 75-microns-sized particles of Al2O3. J Biomed Mater Res. 1996;30:251-260.
- Piattelli M, Scarano A, Paolantonio M, et al. Bone response to machined and resorbable blast material titanium implants: an experimental study in rabbits. J Oral Implantol. 2002;28:2-8.
- Wennerberg A, Albrektsson T, Johansson C, Andersson B. Experimental study of turned and grit-blasted screws shape implants with special emphasis on effects of blasting in material and surface topography. Biomaterials. 1996;17:15-22.
- Mohseni E, Zalnezhad E, Bushroa AR. Comparative investigation on the adhesion of hydroxyapatite coating on Ti-6Al-4V implant: a review paper. Int J Adhes Adhes. 2014;48:238-257.
- Ong JL, Chan D, Chan DCN. Hydroxyapatite and their use as coatings in dental implants: a review. Crit Rev Biomed Eng. 2000;28:1-41.
- Verheyen CCPM, Dhert WJA, Petit PLC, et al. In vitro study on the integrity of a hydroxylapatite coating when challenged with staphylococci. J Biomed Mater Res. 1993:27(6):775-781.
- Ichikawa T, Hirota K, Kanitani H, et al. Rapid bone resorption adjacent to hydroxyapatite-coated implants. J Oral Implantol 1996;22:232-235.
- Johnson BW. HA-coated dental implants: long-term consequences. J Calif Dent Assoc. 1992;20:33-41.
- Rupp F, Liang L, Geis-Gerstorfer J, et al. Surface characteristics of dental implants: a review. Dent Mater. 2018;34:40-57.
- Choi W, Heo SJ, Koak JY, et al. Biological responses of anodized titanium implants under different current voltages. J Oral Rehabil. 2006;33:889-897.
- Teughels W, Van Assche N, Sliepen I, Quirynen M. Effect of material characteristics and/or surface topography on biofilm development. Clin Oral Implants Res. 2006:1768-1781.
- Han A, Tsoi JKH, Rodrigues FP, et al. Bacterial adhesion mechanisms on dental implant surfaces and the influencing factors. Int J Adhes Adhes. 2016;69:58-71.
- Wennerberg A, Albrektsson T, Andersson B. Bone tissue response to titanium implants blasted with fine and coarse particles of aluminum oxide. Int J Oral Maxillofac Implants. 1996;11:38-45.
- Wieland M, Chehroudi B, Textor M, Brunette DM. Use of Ti-coated replicas to investigate the effects on fibroblast shape of surfaces with varying roughness and constant chemical composition. J Biomed Mater Res. 2002;60:434-444.
- Mendonça G, Mendonça DBS, Aragão FJL, Cooper LF. Advancing dental implant surface technology - From micron- to nanotopography. Biomaterials 2008:29:3822-3835.
- Sittig C, Textor M, Spencer ND, et al. Surface characterization of implant materials c.p. Ti, Ti-6Al-7Nb and Ti-6Al-4V with different pretreatments. J Mater Sci Mater Med. 1999;10:35-46.
- Gittens RA, Mclachlan T, Cai Y, et al. The effects of combined micron-/submicron-scale surface roughness and nanoscale features on cell proliferation and differentiation. Biomaterials. 2011;32:3395-3403.
- Brett PM, Harle J, Salih V, et al. Tonetti, Roughness response genes in osteoblasts. Bone. 2004;35:124-133.
- Albertini M, Fernandez-Yague M, Lázaro P, et al. Advances in surfaces and osseointegration in implantology. Biomimetic surfaces. Med Oral Patol Oral Cir Bucal. 2015;20:316-341.
- Zekiy AO. Molecular approaches to functionalization of dental implant surfaces. Eur J Mol Biotechnol. 2015;10:228-240.
- Junker R, Dimakis A, Thoneick M, Jansen JA. Effects of implant surface coatings and composition on bone integration: a systematic review. Clin Oral Implants Res. 2009;20:185-206.
- Takeuchi M, Abe Y, Yoshida Y, et al. Acid pretreatment of titanium implants. Biomaterials. 2003;24:1821-1827.
- Mendonça G, Mendonça DBS, Aragão FJL, Cooper LF. Advancing dental implant surface technology - From micron- to nanotopography. Biomaterials 2008:29:3822-3835.
- Gaggl A, Schultes G, Mu D, Ka H. Scanning electron microscopical analysis of laser-treated titanium implant surfaces - a comparative study. Biomaterials. 2000;21:1067-1073.
- Hallgren C, Reimers H, Chakarov D, et al. An in vivo study of bone response to implants topographically modified by laser micromachining. Biomaterials. 2003;24(5):701-710.
- Frenkel SR, Simon J, Alexander H, et al. Osseointegration on metallic implant surfaces: effects of microgeometry and growth factor treatment. J Biomed Mater Res. 2002;63:706-713.
- Sittig C, Textor M, Spencer ND, et al. Surface characterization of implant materials c.p. Ti, Ti-6Al-7Nb and Ti-6Al-4V with different pretreatments. J Mater Sci Mater Med. 1999;10:35-46.
- Yang Y, Kim KH, Ong JL. A review on calcium phosphate coatings produced using a sputtering process plasma spraying. Biomaterials. 2005;26:327-337.
- de Jonge LT, Leeuwenburgh SCG, Wolke JGC, Jansen JA. Organic-inorganic surface modifications for titanium implant surfaces. Pharm Res. 2008;25:2357-2369.
- de Jonge LT, Leeuwenburgh SC, van den Beucken JJ, et al. The osteogenic effect of electrosprayed nanoscale collagen/calcium phosphate coatings on titanium. Biomaterials. 2010;31:2461-2469.
- Scarano A, Piattelli A, Polimeni A, et al. Bacterial adhesion on commercially pure titanium and anatase-coated titanium healing screws: an in vivo human study. J Periodontol. 2010;81:1466-1471.
- Davies JE. Understanding peri-implant endosseous healing. J Dent Educ. 2003;67(8):932-949.
- Ellingsen JE, Lyngstadaas SP. Bio-Implant Interface: Improving Biomaterials and Tissue Reactions. CRC Press; 2003.
- Bürgers R, Gerlach T, Hahnel S, et al. In vivo and in vitro biofilm formation on two different titanium implant surfaces. Clin Oral Implants Res. 2010;21:156-164.
- Quirynen M, Bollen CM. The influence of surface roughness and surface-free energy on supra- and subgingival plaque formation in man. a review of the literature. J Clin Periodontol. 1995;22:1-14.
- Teughels W, Van Assche N, Sliepen I, Quirynen M. Effect of material characteristics and/or surface topography on biofilm development. Clin Oral Implants Res. 2006:1768-1781.
- Rosenberg ES, Torosian JP, Slots J. Microbial differences in 2 clinically distinct types of failures of osseointegrated implants. Clin Oral Implants Res. 1991;2(3):135-144.
- Rams TE, Roberts TW, Feik D, et al. Clinical and microbiological findings on newly inserted hydroxyapatite-coated and pure titanium human dental implants. Clin Oral Implants Res. (n.d.):2:121-127.
- Zablotsky M, Meffert R, Mills O, et al. The macroscopic, microscopic and spectrometric effects of various chemotherapeutic agents on the plasma-sprayed hydroxyapatite-coated implant surface. Clin Oral Implants Res. 1992;3:189-198.
- Takeshita F, Kuroki H, Yamasaki A, Suetsugu T. Histopathologic observation of seven removed endosseous dental implants. Int J Oral Maxillofac Implants. (n.d.):10:367--372.
- Takeshita F, Ayukawa Y, Iyama S, et al. A histologic evaluation of retrieved hydroxyapatite-coated blade-form implants using scanning electron, light, and confocal laser scanning microscopies. J Periodontol. 1996;67:1034-1040.
- King WJ, Krebsbach PH. Growth factor delivery: how surface interactions modulate release in vitro and in vivo. Adv Drug Deliv Rev. 2012;64(12):1239-1256.
- Petrie TA, Reyes CD, Burns KL, García AJ. Simple application of fibronectin-mimetic coating enhances osseointegration of titanium implants. J Cell Mol Med. 2009;13:2602-2612.
- Gao X, Zhang X, Song J, et al. Osteoinductive peptide-functionalized nanofibers with highly ordered structure as biomimetic scaffolds for bone tissue engineering. Int J Nanomedicine. 2015:10:7109-7128.
Disclaimer from Guident
The views and opinions expressed in articles published in Guident are solely those of the authors and do not necessarily reflect the views of the editors, the editorial board or the publisher, Ivory India. Guident does not endorse any product, technique or treatment described. Clinical content is intended for qualified dental and medical professionals and is not a substitute for individual clinical judgement. Guident, its editors and the publisher accept no liability for any consequence arising from the use of information published in the journal.
How to cite this article
Dr. G. Sreeja, Dr. D. Chalapathi Rao, Dr. G. Harilal. Developments in Dental Implant Surface Modifications – a Review. Guident. 2026 Feb;19(3). Available from: https://guident.net/articles/implantology/developments-in-dental-implant-surface-modifications-a-review.html